1.1 Overview
Using a compression pump to generate compressed air as a driving force to atomize the medicinal solution.
Scope of application: Atomizing drugs through compressed gas and delivering them to the respiratory tract for respiratory useNebulized medication inhalation therapy.
1.2 Product Features and Indicators
1. Input voltage: AC220V, frequency: 50Hz
2. Gas flow rate: Free air flow rate of compression pump ≥ 10L/min
3. Pressure range: Under normal working conditions, the pressure range generated by the atomizer is 80KPa to 150KPa (not specified)When an abnormal situation occurs in the atomizer, the pressure range generated by the atomizer is between 150KPa and400KPa and no occurrence of pipe rupture.
3. Spray rate: * * Atomization rate ≥ 0.2mL/min
4. Residual amount of medication: ≤ 1.0mL
5. Noise: ≤ 60dB (A)
6. Continuous working time: The atomizer should work continuously for more than 4 hours under normal working conditions (timely during this period)Add 0.9% physiological saline to the atomizer cup to ensure normal mist production, and the atomizer should be able to function properly.
7. Equivalent volume particle size distribution: median diameter of fog particles (MMD): 2.7 µ m ± 25%, fog particles with a diameter less than 5 μ mGrain proportion>65%
8. Net weight: approximately 2.1kg
9. Dimensions: 219mm x 210mm x 128mm (length x width x height)
10. Input power: 200VA
11. Safety classification: Do not mix flammable anesthetic gases with air or flammable gases mixed with oxygen or nitrous oxideEquipment used under anesthesia.
12. Transportation and storage environment restrictions:
Environmental temperature range: -20 ℃~+55 ℃
Relative humidity range: 10%~93%
Atmospheric pressure range: 70kPa~106kPa
13. Electrical safety requirements: Class II equipment, Type B application part
14. Immersion protection level: IPX0
16. Normal working conditions:
Environmental temperature range:+5 ℃~+40 ℃
Relative humidity range: 30%~75%