Product Introduction
The infant incubator adopts the "convective heat regulation" method, using microcomputer control technology to provide premature infants with an excellent environment with suitable temperature and humidity, similar to the mother's uterus
The environment,Provide guarantees for the cultivation and care of infants.
Product Features
1、★
Temperature control mode: with skin temperature mode and box temperature mode; Dual CPU high-precision servo temperature control;
2、5.7-inch LCD display screen, capable of simultaneously displaying set values and display values, with 24-hour real-time temperature display; Curve traceability; Having information storage and retrieval functions;
3、★
Real time display of heating power percentage
4. Equipped with humidity display and control functions, oxygen concentration display and control functions;
5. The product has a power on self-test function and 24 types of safety and fault alarms;
6. Equipped with an independent second thermal cut-off device for safety protection
7. Spiral shell biomimetic air duct and DC centrifugal fan generate air pressure difference to ensure that fresh air is always sucked in;
8. The overall thermal storage aluminum water tank can significantly reduce temperature fluctuations;
9、★The front panel has a temperature correction function, and the skin temperature sensor has a detachment alarm function.
10. 37 ℃ temperature setting function, double-layer constant temperature cover; Better insulation performance
11. Equipped with an independent locking device for the main entrance;
12. Embedded integrated sensor box, drawer style and heated integrated water tank;
13. Baby crib tilt angle: ± 12 ° infinitely adjustable
14. Optional height adjustment function for the entire machine
15. Optional upper and lower blue light function
Product parameters
1. Box temperature control temperature range: 25.0 ℃~37.0 ℃; 37.1 ℃~39.0 ℃ (crossover mode)
2. Skin temperature control temperature range: 34.0 ℃~37.0 ℃; 37.1 ℃~38.0 ℃ (crossover mode)
3. Temperature display accuracy: 0.1 ℃
4. Temperature variability: not greater than 0.5 ℃
5. Skin temperature sensor accuracy: ± 0.2 ℃
6. Temperature uniformity (horizontal position of mattress): not greater than 0.8 ℃
7. Temperature uniformity (mattress tilt position): not greater than 1.0 ℃
8. Difference between skin temperature sensor temperature and control temperature: not greater than 0.7 ℃
9. The difference between the average incubator temperature and the control temperature shall not exceed 1.5 ℃