FH1800 (W) | W series fume hood

1、 Technical parameters

1. Basic parameters of fume hood:

(1) External dimensions ≥ (L × D × H) 1800mm × 800mm × 2200mm; (Size can be customized and modified according to requirements)

(2) The internal dimensions are ≥ (L × D × H) 1580mm × 550mm × 1670mm.

(3) Inlet wind speed: 0.3-0.8m/s;

(4) System exhaust volume: 1850 m ³/h;

(5) Rated power: 800 W (excluding the power of cabinet socket load (load cannot exceed 500W));

(6) Noise level: ≤ 65dB (A);

(7) Illuminance: ≥ 800lx;

(8) Front window glass opening height: 1300mm;

(9) LED fluorescent lamp power: 16W * 2;

(10) Fan model: Fanshida FH400 fan, speed: 1400 RPM, flow rate: 3650 m ³/h, power: 370W;

(11) Number of users: 2-3 people;

(12) Wooden packaging dimensions: Cabinet: 1940 * 940 * 2410, external exhaust fan 1040 * 890 * 710

2、 Product concept:

In the chemical laboratory, various harmful odorous gases, odors, moisture, and corrosive substances are generated during experimental operations. In order to protect the safety of users and prevent the spread of pollutants from the experiment to the laboratory, a fume hood should be used near the pollution source.

Attention: Strong acid and strong alkali, flammable and explosive experiments, cannot use this type of fume hood.

3、 Structural and functional characteristics:

1. The outer shell of the fume hood is made of 1.0mm cold-rolled steel plate, which has undergone rust prevention treatment and electrostatic spraying, and has good smoothness and corrosion resistance.

2. The operation area of the fume hood adopts anti duplication special board, which has high acid and alkali resistance;

3. The front window of the fume hood is a manual window that can stop at any height within the travel range.

4. The front window glass of the fume hood is 5mm thick tempered glass.

5. The control panel of the fume hood adopts a ship type switch to control the power supply and fluorescent lamps of the cabinet, and uses a wind speed adjustment knob to adjust the fan gear, achieving stepless speed regulation. The machine has a beautiful appearance and is easy to operate.

6. The electrical control system of the fume hood has functions such as overload prevention and electric shock prevention, with stable performance and long service life.

7. The ventilation cabinet socket adopts laboratory specific safety products with transparent splash proof covers, which are beautiful and durable.

8. The fume hood is equipped with a wall mounted water supply cock, water supply flow control valve, wall mounted gas supply cock, gas supply flow control valve, and laboratory sink for easy experimentation.

9. External AC backward centrifugal fan for fume hood, with high air volume and low noise.

10. The fume hood has a power-off memory function, which means that in the event of a sudden power outage, it can be powered on again to maintain the working state before the power outage, facilitating experimental operations.