Product Introduction
The neuromuscular electrical stimulator is suitable for conducting neuromuscular training through patient surface electrical stimulation.
Product parameters
1. Operating frequency of the product
The fundamental frequency of the output pulse in Mode I is 500Hz, with a tolerance of ± 10%; The pulse modulation frequency can be adjusted within the range of 0.5Hz to 5Hz.
The pulse frequency of Mode II is adjustable within the range of 0.5Hz to 5Hz, with a tolerance of ± 10%.
2. Pulse width
Mode I outputs a pulse width of 1ms with a tolerance of ± 30%.
The output pulse width of Mode II is 10ms, with a tolerance of ± 30%.
3. Modulation wave pulse width
The modulation wave width is 10ms, with a tolerance of ± 30%.
4. Treatment time
Set the range from 5 minutes to 30 minutes, with a tolerance of ± 5%.
5. Output current
Each group of the stimulator outputs independently. When the load impedance is 1k Ω, the output current of each group can be continuously adjusted from 0 to 99 levels, with an output value of 100mA and a tolerance of ± 30%.
6. Output waveform
The output carrier waveforms of Mode I and Mode II are bidirectional asymmetric waves.
7. Continuous working hours
The equipment should work for no less than 4 hours at * * power.