ANP96 | Nucleic acid extractor

working principle

This instrument is a high-tech product that uses the universal magnetic bead method to extract nucleic acids. It has the advantages of high automation, fast extraction speed, stable results, and easy operation. Using a dedicated 96 hole deep hole plate, 1-96 samples can be operated simultaneously.

Using the magnetic rod on the experimental cabin magnetic rod rack, move the magnetic beads adsorbed with nucleic acid to different reagent wells, and then use the stirring sleeve on the outer layer of the magnetic rod to repeatedly and rapidly stir the liquid, so that the liquid and magnetic beads are evenly mixed. After cell lysis, nucleic acid adsorption, cleaning and desorption, high-purity nucleic acid is finally obtained.

Scope of application

Used for extracting and purifying nucleic acids from clinical samples.

Performance characteristics

Screen operation

10.1-inch touch screen in Chinese, simple and easy to use

Precise control

The interface is simple and easy to understand, with detailed instructions, real-time display of program running status, and can be used independently without the need for a computer

temperature control

Customize cracking and elution temperatures according to requirements, with room temperature ranging from+5 ℃ to 120 ℃ freely adjustable

Temperature maintenance

The sample lysis module and elution module have temperature control and maintenance functions

Free programming

Powerful program editing function, open for editing all reagents to meet different reagent needs

Quick Extraction

Short operation time (depending on the selected reagent), high throughput, capable of extracting 1-96 samples simultaneously

Magnetic rod module

Integrated magnetic rod, magnetic flux 4800-5200GS

extraction efficiency

Purification schemes can be optimized based on different reagents to achieve higher extraction efficiency and purity

Stable and reliable

Fully automated completion, avoiding differences and errors caused by manual operation, stable results, and good repeatability

Pollution prevention and control

Built in UV disinfection function, capable of timed disinfection, and equipped with a ventilation port to avoid cross contamination

Safe and reliable

Pause upon opening the door to prevent injury to the equipment during operation

Reagent open

Various manufacturers' reagents can be used for extraction

Scan code gun (optional)

Equipped with a barcode scanner, reagents are individually numbered for easy later reference

Program import

Support automatic program import, multiple devices can be edited only once

Low temperature refrigeration

Equipped with low-temperature refrigeration function, extending the effective time of reagents


Work environment requirements

The working environment of the nucleic acid extractor should meet the following requirements:

a) Only applicable indoors

b) Environmental temperature: 10 ℃~30 ℃;

c) Relative humidity: ≤ 70%;

d) Atmospheric pressure range: 86kPa to 106kPa;

e) There is no strong vibration or corrosive gas present around;

f) Avoid direct sunlight exposure or the influence of other cold and heat sources;

g) Stay away from strong electromagnetic field interference;

h) Power supply voltage: AC220V, 50Hz;

technical parameter


model

ANP96

Extraction method

Magnetic bead extraction method

Sample flux

1~96T

Equipment size

780*380*565(L*W*H)

Instrument weight

51KG

Processing volume

60~1000μL

Magnetic flux of magnetic rod

4800~5200Gs

operating system

Linux system

Operation interface

10.1-inch full Chinese LCD touch screen

Heating method

Heating film

Low temperature refrigeration

Semiconductor refrigeration

Program import

USB automatic import

Scanning gun

Optional selection

Work noise

<65dB

Oscillation mode

Up and down oscillation with three adjustable gears

Extraction time

1560min

UV sunlight

have

Magnetic bead collection efficiency

≥98%

Use power supply

AC220V,50Hz

Equipment expiration date

8 years