working principle
This instrument is a high-tech product that uses the universal magnetic bead method to extract nucleic acids. It has the advantages of high automation, fast extraction speed, stable results, and easy operation. Using a dedicated 96 hole deep hole plate, 1-32 samples can be operated simultaneously.
Using the magnetic rod on the experimental cabin magnetic rod rack, move the magnetic beads adsorbed with nucleic acid to different reagent wells, and then use the stirring sleeve on the outer layer of the magnetic rod to repeatedly and rapidly stir the liquid, so that the liquid and magnetic beads are evenly mixed. After cell lysis, nucleic acid adsorption, cleaning and desorption, high-purity nucleic acid is finally obtained.
Scope of application
Used for extracting and purifying nucleic acids from clinical samples.
Performance characteristics
Screen operation | 7-inch resistive touch screen in Chinese, simple and easy to use |
Precise control | The interface is simple and easy to understand, with detailed instructions, real-time display of program running status, and can be used independently without the need for a computer |
temperature control | Customize cracking and elution temperatures according to requirements, with room temperature ranging from+5 ℃ to 120 ℃ freely adjustable |
Temperature maintenance | The sample lysis module and elution module have temperature control and maintenance functions |
Free programming | Powerful program editing function, open for editing all reagents to meet different reagent needs |
Quick Extraction | Short operation time (depending on the selected reagent), high flux, capable of extracting 1-32 samples simultaneously |
Magnetic rod module | Integrated magnetic rod, magnetic flux 4800-5200GS |
extraction efficiency | Purification schemes can be optimized based on different reagents to achieve higher extraction efficiency and purity |
Stable and reliable | Fully automated completion, avoiding differences and errors caused by manual operation, stable results, and good repeatability |
Pollution prevention and control | Built in UV disinfection function, capable of timed disinfection, and equipped with a ventilation port to avoid cross contamination |
Safe and reliable | Pause upon opening the door to prevent injury to the equipment during operation |
Reagent open | Various manufacturers' reagents can be used for extraction |
Work environment requirements
The working environment of the nucleic acid extractor should meet the following requirements:
a) Only applicable indoors
b) Environmental temperature: 10 ℃~30 ℃;
c) Relative humidity: ≤ 70%;
d) Atmospheric pressure range: 86kPa to 106kPa;
e) There is no strong vibration or corrosive gas present around;
f) Avoid direct sunlight exposure or the influence of other cold and heat sources;
g) Stay away from strong electromagnetic field interference;
h) Power supply voltage: AC220V, 50Hz;
technical parameter
model | ANP32 |
Sample flux | 1-32T |
Magnetic rod (fixed) | 32 pieces |
Processing volume | 60~1000μL |
Magnetic bead collection efficiency | ≥98% |
The most suitable size of magnetic beads | 0.2~1.0μm |
Operation interface | Touch control operation |
ultraviolet radiation | have |
Types of orifice plates used | 96 hole deep hole plate |
Use power supply | AC220V,50Hz |
Instrument weight | 23KG |
Overall size (mm) | 420*358*475(L*W*H) |
Equipment expiration date | 8 years |