ANP32 | Nucleic acid extractor

working principle

This instrument is a high-tech product that uses the universal magnetic bead method to extract nucleic acids. It has the advantages of high automation, fast extraction speed, stable results, and easy operation. Using a dedicated 96 hole deep hole plate, 1-32 samples can be operated simultaneously.

Using the magnetic rod on the experimental cabin magnetic rod rack, move the magnetic beads adsorbed with nucleic acid to different reagent wells, and then use the stirring sleeve on the outer layer of the magnetic rod to repeatedly and rapidly stir the liquid, so that the liquid and magnetic beads are evenly mixed. After cell lysis, nucleic acid adsorption, cleaning and desorption, high-purity nucleic acid is finally obtained.

Scope of application

Used for extracting and purifying nucleic acids from clinical samples.

Performance characteristics

Screen operation

7-inch resistive touch screen in Chinese, simple and easy to use

Precise control

The interface is simple and easy to understand, with detailed instructions, real-time display of program running status, and can be used independently without the need for a computer

temperature control

Customize cracking and elution temperatures according to requirements, with room temperature ranging from+5 ℃ to 120 ℃ freely adjustable

Temperature maintenance

The sample lysis module and elution module have temperature control and maintenance functions

Free programming

Powerful program editing function, open for editing all reagents to meet different reagent needs

Quick Extraction

Short operation time (depending on the selected reagent), high flux, capable of extracting 1-32 samples simultaneously

Magnetic rod module

Integrated magnetic rod, magnetic flux 4800-5200GS

extraction efficiency

Purification schemes can be optimized based on different reagents to achieve higher extraction efficiency and purity

Stable and reliable

Fully automated completion, avoiding differences and errors caused by manual operation, stable results, and good repeatability

Pollution prevention and control

Built in UV disinfection function, capable of timed disinfection, and equipped with a ventilation port to avoid cross contamination

Safe and reliable

Pause upon opening the door to prevent injury to the equipment during operation

Reagent open

Various manufacturers' reagents can be used for extraction

Work environment requirements

The working environment of the nucleic acid extractor should meet the following requirements:

a) Only applicable indoors

b) Environmental temperature: 10 ℃~30 ℃;

c) Relative humidity: ≤ 70%;

d) Atmospheric pressure range: 86kPa to 106kPa;

e) There is no strong vibration or corrosive gas present around;

f) Avoid direct sunlight exposure or the influence of other cold and heat sources;

g) Stay away from strong electromagnetic field interference;

h) Power supply voltage: AC220V, 50Hz;

technical parameter

model

ANP32

Sample flux

1-32T

Magnetic rod (fixed)

32 pieces

Processing volume

60~1000μL

Magnetic bead collection efficiency

≥98%

The most suitable size of magnetic beads

0.2~1.0μm

Operation interface

Touch control operation

ultraviolet radiation

have

Types of orifice plates used

96 hole deep hole plate

Use power supply

AC220V,50Hz

Instrument weight

23KG

Overall size (mm)

420*358*475(L*W*H)

Equipment expiration date

8 years